Electrical Mapping
Author: Fernando Campos fernando.campos@kcl.ac.uk
This tutorial demonstrates how to compute local activation times (LATs) and action potential durations (APDs) of cells in a cardiac tissue.
Electrical Mapping
The wavelength of the cardiac impulse, given by the product of conduction velocity (CV) and refractory period, is of utmost importance for the study of arrhythmia mechanisms. The assessment of the CV as well as the refractory period of the cardiac action potential (AP) relies on the determination of activation and repolarization times, respectively. Experimentally, these are usually calculated based on 1) transmembrane voltages \(V_{\mathrm m}\) measured by glass micro-electrodes or in optical mapping; or 2) extracellular potentials \(\phi_{\mathrm e}\) measured at the surface of the tissue.
The estimation of activation/repolarization times is based on an event detector that records the instants when the selected event type occurs. Local activation time (LAT), for instance, is usually taken as the time of maximum \(dV_{\mathrm m}/dt\) or the minimum \(d\phi_{\mathrm e}/dt\):
Instead of using derivatives, LATs can also been estimated by the crossing of a threshold value provided that the chosen value reflects the sodium channel activation since they are responsible for the rising phase of AP. One could, for instance, pick the time instant when \(V_{\mathrm m}\) crosses -10 mV with a positive slope. Equally, one could look for crossing of \(V_{\mathrm m}\) with a negative slope to detect repolarization events. AP duration (APD) can then be computed as the difference between repolarization and activation times. The detection of activation (\(t_{\mathrm act}\)) and repolarization (\(t_{\mathrm rep}\)) times based on threshold crossing is illustrated below:
Problem Setup
This example will run a simulation using a 2D sheet model (1 cm x 1 cm). A transmembrane stimulus current of 100 uA/cm^2 is applied for a period of 1 ms at the middle of the left side of the tissue. A ground electrode is placed at the opposite side of the stimulating electrode. Simulation of electrical activity is based on the bidomain equations. Cellular dynamics cardiac myocytes within the tissue are described using the Modified Beeler-Router Drouhard-Roberge (MBRDR) model. Activation times are obtained either from \(V_{\mathrm m}\) or \(\phi_{\mathrm e}\) using a threshold crossing method, whereas repolarization can only be obtained by threshold crossing from \(V_{\mathrm m}\). Estimation of repolarization times from \(\phi_{\mathrm e}\) is out of the scope of this tutorial.
Usage
To run this tutorial :
cd
tutorials/08_lats
./run.py
--help
The following optional arguments are available (default values are indicated):
--quantity Quantity/variable/signal
used to compute activation time
Options: {vm,phie}
Default: vm
--act-threshold The
threshold value for determining activation from :math:`V_{\mathrm m}` or :math:`\phi_{\mathrm e}`
Default: -10 mV
--rep-threshold The
threshold value for determining repolarization from :math:`V_{\mathrm m}
Default: -70 mV
--show-APDs Visualize
APDs instead of LATs
If the program is run with the --visualize
option, meshalyzer will
automatically load the computed activation sequence (LATs). If the --show_APDs
is given, then the computed APDs will be loaded:
Visualize
LATs:
./run.py
--visualize
Visualize
APDs:
./run.py
--show-APDs
Interpreting Results
Differences between activation sequences computed from \(V_{\mathrm
m}\) or \(\phi_{\mathrm e}\) and/or with
different thresholds can be appreciated with meshalyzer (--visualize
).
The activation sequence for the default parameters looks like this:
Note
- It is important to note that while computing activation from \(V_{\mathrm m}\) -lats[].mode has to be set to zero. This is because \(V_{\mathrm m}\) is about -85 mV in cardiac cells at rest and rapidly rises to about +40 mV during the activation phase. This means \(V_{\mathrm m}\) will cross the threshold with a positive slope. While computing repolarization, on the other hand, \(V_{\mathrm m}\) goes from positive values back to the rest. The threshold will be then crossed with a negative slope.
Note
- Unlike the shape of \(V_{\mathrm m}\) which remains pretty much the same in all cells, \(\phi_{\mathrm e}\) morphology may vary depending on its position in relation to both stimulation and ground electrodes. Thus, LATs obtained based on \(\phi_{\mathrm e}\) might differ from those obtained from \(V_{\mathrm m}\).
- If the criteria for threshold crossing detection is never satisfied, the resulting entry in the output file will be -1. See action sequence obtained from \(\phi_{\mathrm e}\).
The computed APDs using the default parameters looks like this (use --show-APDs
instead of --visualize
):
CARPentry Parameters
The relevant parameters used in the tissue-scale simulation are shown below:
#
Number of events to detect
-num_LATs =
2
# Event
1: activation
-lats[0].ID =
'ACTs'
-lats[0].all =
0
-lats[0].measurand =
0
-lats[0].threshold =
-10
-lats[0].mode =
0
# Event
2: repolarization
-lats[1].ID =
'REPs'
-lats[1].all =
0
-lats[1].measurand =
0
-lats[1].threshold =
-70
-lats[1].mode =
1
num_LATs referes to the number of events we want to detect. In this tutorial there are two events: activation and repolarization. -lats[0] are options to compute activation while -lats[1] are options to compute repolarization. -lats[].ID is the name of output file; -lats[].all tells CARP if it should detect only the first event (-lats[].all = 0), i.e. first beat, of all of them (-lats[].all = 1); -lats[].measurand refers to the quantity: \(V_{\mathrm m}\) (0) or \(\phi_{\mathrm e}\) (1); -lats[].threshold is the threshold value for determining the event; finally, -lats[].mode tells CARP to look for threshold crossing with a either a positive or negative slope.
'